UCSF Titan Krios 1 data collection settings

Nominal Magnification

Pixel Size (Å/pix)

Super Resolution

Pixel Size (Å/pix)

Counting mode (or after 2x FT bin)

Dose Rate 

(e-/pix/s); 

Total Dose (e-/A2)

Exposure time (s)

Dose (e-/A2/frame)

Magnification of 5 μm physical pixel (super res)

64,000

0.685 1.371

12.0 ; 63.9

10

0.54 (0.085s frame)

36496 (72939)
105,000 0.4094 0.8189 16 ; 47.7 2 0.596 (0.025s frame) 61057 (122115)

130,000

0.332

0.664

16 ; 47.2

1.3

0.54 (0.015s frame)

75188 (150602)

 

  • Approximate single particle data collection speeds at 105kx (actual speeds vary based on size of image shift groups after combining targets)
    • 5x5 hole IS group (1.2/1.3um foil), physical pixel size: ~11k/24hr
    • 5x5 hole IS group (1.2/1.3um foil), super res pixel size: ~8k/24hr
    • 3x3 hole IS group (1.2/1.3um foil), physical pixel size: ~8k/24hr
    • 3x3 hole IS group (1.2/1.3um foil), super res pixel size: ~6k/24hr
  • Use iLab to reserve Krios1 time (click here for instructions) (link to iLab)
  • If another magnification or dose rate necessary please contact [email protected]
  • UCSF MotionCorr2 v1.3.0 will be used for real time motion correction and dose-weighting.
  • Other useful protocols